- Respiratory Rate and Minute
Ventilation Volume can be independently regulated at random to meet various patients'
requirements.
- Manual
Ventilation Mode, PEEP
- The
airway pressure can be shown directly and reduced automatically.
- Air-Oxygen
Mixer
- Low-pressure
Alarm
- Assistant
Ventilation Mode can support oxygen for patients’ inspiratory
effort by automatic
flow trigger
Standard
configuration
Ventilator ( including air-oxygen mixer, automatic
airway pressure relief valve ), small oxygen cylinder, portable bag,
laryngoscope, phlegm suction pot, humidifier, airway opener, extension line,
oxygen refill fitting, wrench, face mask, nasal oxygen canula, endotracheal tube.
| Physical
parameters |
|
| Weight
of Main Device |
0.70kg |
| Volume
of Main Device |
178×46×68mm |
| Weight
of Gas Cylinder |
2.5kg |
| Volume
of Gas Cylinder |
2L |
|
Main
technical indicator |
|
| Breathing
Mode | IPPV
Mode; A/C Ventilation; Manual Ventilation; PEEP |
|
Operating
Pressure |
0.35--0.40
MPa |
|
Minute
Ventilation Volume |
4--25L/min(Adjustable
in succession) |
|
Respiratory
Rate | 10-60
times/min (Adjustable in succession) Seven Indication: 10,15,20,30,40,50,60
times/min |
|
I /E
Ratio | 1:2
|
|
Oxygen
Concentration |
≤60% |
| Inspiratory
Trigger |
≤-1Kpa
(±50Pa) |
|
Low Pressure
Alarm | ≤0.25
MPa |
|
High
Pressure Pause |
0.9-1.1
MPa |
|
Reverse
Flow Pressure of Airway |
≤6
KPa |
|
Noise
of Main Device |
65dB |
| |
|
| Supply
Flow of Gas Cylinder |
60 L/min |
| Continuous
Working Time of Cylinder |
Around
90min |
|
Oxygen
Storage Pressure for Cylinder |
14.7
MPa |
|
Oxygen
Consumption |
Adult
≤ 2 Cylinders/24h Cylinder Specification: 12250 Kpa/40L |
|
Protective
Function |
|
| Airway
Pressure Display |
Visual |
| Low
Pressure Alarm |
Warning
when supplied air pressure drops below 2.6bar |
|
Airway
Pressure Release Valve |
Reducing
pressure automatically when airway pressure exceeds 50cmH2O |
| High
Pressure Protection |
When
supplied air pressure is over the working pressure, the ventilator will stop
working following audible alarm voice. |
(Respiratory
Rate, Minute Ventilation Volume, I:E Ratio, Oxygen Concentration, Inspiratory
Trigger are subject to a tolerance of ±15%)