| FHR |
- Dual pulsed
Doppler
- · Intensity:
10mW/cm²
- Accuracy:
±2% of range
| - Ultrasound
frequency: 0.985MHz
- FHR
range : 30-240 bpm FHR
|
|
FM | - Auto
detection of dual fetal movement
| |
| UC | - External
Type
- Reference
(zero) control
| - Frequency
response: DC~0.5HZ
- Measurement
range: 0~99units
|
|
Printer | - Thermal
array type
- Auto
print: off, 10,20,30,40,50,60,min
| - Print
speed: 1,2,6cm/min, high speed
- Paper
feeding function
|
|
Display |
- 3 channels
(FHR I, FHR II, UC)
- 7"
Wide TFT-color LCD: LCD type
- Large
7-segment & volume level LED: LED type
| |
| Sound | - Doppler
sound
- Information
Sound
| |
| Set-up | - Alarm
upper/Lower value
- Print
Speed
| |
|
Function | - Mark
function
- Auto
print function
- Multi-language
& quick guide function: LCD type
| - FHR
II offset function
- Fetal
movement print function
|
|
Trend | - Data
saving for 450 hours (3hr/person): LCD type
| |
| Power |
- Input: AC
100-240V(50/60Hz)
- Battery
operation
| - Output:
DC 18V, 2.5A
- Consumption:
80VA
|
|
Dimension | - 80mm(H)
x 330mm(L) x 280mm (D)
| |
| External
Link | | |
|
Accessory
standard | - Ultrasound
Doppler probe 2ea
- Event
marker jack 1ea
- Power
adaptor & Cord 1ea
- Probe
belt 3ea
| - UC
probe 1ea
- Probe
holder 2ea
- Probe
paper 2ea
- Ultrasound
gel 1ea
- Operation
manual 1ea
|
|
Optional | - AST
(Acoustic stimulator)
- Bluethe
system
| - Rechargeable
battery
- Wall