Multipara
Monitor MAX-12

In-hospital
applications include emergency pre-and post-operative
care, ambulatory surgery, intermediate care/step
down units, labor and delivery, and hospital-based
special procedure areas. Suitable for use in physicians’
offices, clinics, outpatient surgical centers, extended
care facilities and other patient care areas, which
of require affordable monitoring
- 12.1"
wide TFT LCD screen with maximum 11-waveform
display. Resolution 800 x600.
- Advanced
streamline outline design, portable, compact,
lightweight
- SpO2
Pulse Tone Modulation
- Easy
operation with user friendly menu structure
design and rotary dial
- Protection
Against electrosurgical interference and defibrillation
and no need to disconnect the monitor from the
patient in process of defibrillating.
- Intelligent
audio and visual comprehensive alarm
- Rechargeable
high-energy built-in Li-Ion battery
- 96
hour graphical and tabular trends memory
- Can
be used to monitor Adult, Paed, Neonates
- Upto
7-Channel ECG display, Big Character Display,
OxyCRG, Trend Table Views.
- With
function of Arrhythmia and ST segment analysis,
drug dose calculation, pacemaker detection
- Memory
storage of upto 500 NIBP measurements, 60-alarm
events storage with 120-seconds full disclosure
of ECG
Standard
configuration: ECG, HR, PR, SpO2, ST analysis,
Arrhythmia Detection, RESP (RA-LL impedance), NIBP,
TEMP*2(surface and rectal type), Drug dose calculation.
Optional
configuration: Recorder and 2-IBP, EtCO2
Technical
Specifications:
|
ECG
Patient Type: Adult/Paed/Neonate
Lead: 5-Lead : RA, LA, RL, LL, V
Lead Selection: I, II, III, aVR, aVL, aVF,
V
ECG Waveforms: 2 or 6 or 7 waveforms
Gain Selection: x2.5mm/mV, x5.0mm/mV,
x10mm/mV,
20mm/mV or auto
Sweep Speed: 12.5mm/s, 25mm/s, 50mm/s
Heart
Rate Range: Adult: 15~300BPM
Paed or Neonate: 15~350BPM
Resolution: 1 BPM
Accuracy: ± 1 % or ± 1 BPM, Whichever is higher
Operation Modes: Diagnostic, Monitoring and
Surgery
S-T
Segment: YES
Arrhythmia Analysis: YES
Alarms:
Yes, Audio-visual, Alarm events recallable
Protection:
Withstand 4000VAC/50Hz voltage in
isolation against Electrosurgical
interference and Defibrillation
|
Respiration
Method:
Impedance between RA-LL
Measurement
Range: Adult: 0-120rpm
Paed/Neonate: 0-150rpm
Resolution:
1rpm
Accuracy:
± 2 rpm
Alarm
Range: Adult: 0-120 rpm
Paed/Neonate: 0-150 rpm
Apnea
Alarm: YES
Apnea Delay: 10-40 sec
|
|
CO2
(Optional)
Measurement
range: 0~99 mmHg
Accuracy:
± 2 mm Hg (0~40 mmHg) Or ±10 % (77~99 mmHg)
|
|
NIBP
Method:
Oscillometric
Operation:
Mode: Automatic/Manual/Continuous
Measurment
Modes: Systolic/Diastolic/Mean Pressure
Measurement
Range: Adult: 10-270mmHg
Paed: 10-200mmHg
Neonate: 10-135mmHg
Overpressure
Protection: YES
Resolution: 1mmHg
Alarms:
Systolic/Diastolic/Mean
|
Temperature:
Units:
°C or °F
Measurement
Range: 0-50 °C
Resolution:
0.1 °C
Accuracy:
± 0.1 °C (excludes the sensor)
Channel:
2-Channels
Alarm
Range: 0-50 °C
|
|
SpO2
Measurement
Range: 0-100%
Accuracy:
± 2 digits, 70-99%, Adult/Paed
±
3 digits, 70-99%, Neonate
Unspecified, Less than 70%
|
IBP
(Optional)
Measurement
Range: -50 – 300 mmHg
Channel:
2-Channels
Pressure
Sensor: Sensitivity: 5uV/V/mmHg
Impedance: +300~+10000 Ohms
Pressure
Name: ART, PA, CVP, RAP, LAP, ICP
Resolution:
1 mmHg
Accuracy:
± 2 or ± 1 mmHg, whichever is greater
excludes sensor
Alarm
Range: -50 – 300 mmHg
|
|
Pulse
Rate
Measurement
Range: 18-254 BPM
Alarm
Range: 18-254 BPM
Accuracy:
± 3 BPM
Resolution:
1 BPM
|
Power:
100~240 VAC, 50/60 Hz
Battery:
Rechargeable Li-ion, 2A/hr, provides
backup of upto 2 hrs
Dimension: 318 (W) x 152 (D) x 264 (H) mm
Weight:
5 Kg
|

In-hospital applications
include emergency, pre-and post-operative care, ambulatory surgery, intermediate
care/step down units, labor and hospital-based special procedure areas.
- Advanced streamline outline design, compact,
lightweight, in-built handle makes it truly portable
- Can
be used to monitor adult, pediatric & neonate patient
- Upto
eight wave-channel display
- In-built
Rechargeable battery which provides a good backup lime
- 96
Hours graphical and tabular trends memory
- With
function of arrhythmia analysis, drug dose calculation, pacemaker detection and
ST segment analysis.
- User friendly
menu structure with easy to use rotary dial
- Protection
Against electro surgical interference and defibrillation
- Intelligent
audio and visual comprehensive alarm. Volume adjustment possible
- Central
monitoring system for long-distance monitoring, diagnosis, maintenance and software
upgrade possible
Standard
Configuration: ECG, HR, Sp02' ST Analysis, Resp, NIBP, Temp, Drug
Dose Calculation, built in battery Optional Configuration: Recorder,
2*IBP, CO2
Technical
Specifications:
| ECG
Input:
5 wires/3 wires ECG cable Lead
selection: I,II,III, aVR, aVL, aVF, V Gain
(mm/mV) : 1/4,1/2, 1, 2, 4 & auto Frequency response: Diagnosis 0.05 -
100Hz
Monitor 0.5
- 40 Hz
Surgery 1.0 - 25 Hz Sweep speed (mm/sec) : 12.5, 25, 50 Heart rate range:
10-300 BPM Heart
rate accuracy: ± 1 % or ± 1 BPM, which is greater ST segment deviation analysis:
-0.8 - +0.8mV ST
accuracy: ±0.02mV Heart
Rate Alarm: Yes Non-Invasive
Blood Pressure (Oscillographic) Measurement type: adult, pediatric, neonate
Measurement range: Systolic 30~270mmHg
Diastolic 10~220mmHg
Mean 20~235mmHg Accuracy:
± 2% or ± 3mmHg, which is greater Resolution:
1 mmHg Protection:
over pressure Pulse
Oximetry SpO2
measurement range: 0 -100 % Resolution:
1 % Pulse
measurement range: 25- 250BPM Accuracy:
± 2 %. 70~100% Resolution:
1 BPM Physical
Weight:
~ 4.5 Kg Size:
318mm x 264mm x 152mm

OxyCRG
Diagram Enables the doctor to relate information of patient rapidly &
accurately by displaying HR, RESP and oxygen parameter on same screen
 Simultaneous
Seven leads ECG display | Temperature
(Skin or Intra-Cavity) Measurement range: 0~500C Accuracy: ± 0.10C
Resolution:
0.10C Response
time: < 3 min. Channel:
1 Channel Respiration Rate (RA-LL Impedance) Measurement
range: 0 - 120 BPM Accuracy: ± 2 BPM Resolution:
1 BPM EtC02
(Optional Side Stream Type) Measurement range: 0%-13.1 % (0~99mmHg) Accuracy:
± 0.3%(< 5% Measurement value) <10%
(> 5% Measurement value) Response Time: 180 ms 2·lnvasive
Blood Pressure (Optional) Measurement range: -45~+300mmHg Channel:
2 channel Transducer
sensitivity: 5uV/V/mmHg, 2% Unit
display: kPa or mmHg selectable Environment
Operation
Storage
Temperature 0- 40·C -20 - 500C Relative humidity < 85%
< 93% Electrical
Input:
100~240 V AC, 50/60Hz Consumption:
< 70VA Battery
type: Lead-Acid Battery
Operation: ~ 2 hrs 
Big
Digit Display Big digit display format makes reading parameters from
a distance easy |
Next