Multi-Parameter
Patient Monitor
MAX-10 10.4" Color TFT
LCD Screen

In-hospital
applications include emergency, pre-and post-operative care, ambulatory surgery,
and intermediate care/step down units, labor and hospital-based special procedure
areas. -
10.4"
wide TFT LCD screen with maximum 9-waveform display. Resolution 800 x480. -
Can be used to monitor Adult, Paed.
Neonates -
Advanced
streamline outline design, portable, compact, lightweight -
SpO2 Pulse Tone Modulation -
Easy operation with user friendly
menu structure design and rotary dial -
Protection
Against electro surgical interference and defibrillation -
With function of Arrhythmia and
ST segment analysis, drug dose calculation, pacemaker detection -
Intelligent audio and visual comprehensive
alarm -
Rechargeable
high-energy built-in Li-Ion battery -
96
hour graphical and tabular trends memory -
Upto
7-Channel ECG display, Big Character Display, OxyCRG. Trend Table Views -
Memory storage of upto 500 NIBP
measurements, 50-alarm events storage with 120-seconds full disclosure of ECG
Standard
Configuration: ECG, HR, SpO2, ST Analysis, Resp, NIBP, 2*Temp,
Drug Dose Calculation, built in battery
Optional Configuration: Recorder, 2*IBP
Technical Specifications:
| ECG
| |
Patient Type |
: Adult/Paed/Neonate | |
Lead: 5-Lead |
: RA LA. Rl, Ll, V | |
Lead Selection |
: I, II. Ill, aVR, aVl, aVF, V | |
ECG Waveforms |
: 2 or 6 or 7 waveforms | |
Gain Selection |
: x2.5mm!mV, x5.0mm!mV, x10mmlmV, 20mm/mV or auto | |
Sweep Speed |
:12.5mm/s, 25mm/s, 50mm/s | |
Heart Rate Range |
:Adult. 15-300BPM | |
Paed or Neonate |
:15-350BPM | |
Resolution |
: 1 BPM | |
Accuracy |
: ± 1 % or ± 1 BPM, Whichever is higher | |
Operation Modes |
:Diagnostic, Monitoring and Surgery | |
S- T Segment |
:YES | |
Arrhythmia Analysis |
:YES | |
Alarms |
Yes. Audio-visual Alarm event:> recallable | |
Protection |
:Withstand 4000VAC/50Hz voltage in Isolation against Electrosurgical Interference
and Defibrillation | | |
| Non-Invasive
Blood Pressure | |
Method |
: Oscillometric | |
Operation Mode |
: Automatic/Manual/Continuous | |
Measurment Modes |
: Systolic/Diastolic/Mean Pressure | |
Measurement Range: |
Adult: 10-270mmHg Paed . 1o-200mmHg Neonate : 10-135mmHg Overpressure Protection:
Yl:.S | |
Resolution |
1 mmHg | |
Alarms |
Systolic/Diastolic/Mean | | |
| |
| Pulse
Oximetry | |
Measurement Range |
0-100% | |
Accuracy | ±
2 digits. 70-99%. Adult/Paed ± 3 digits. 70-990/. Neonate Unspecified,
Less than 70% |
OxyCRG Diagram
Enables
the doctor to relate Information of patient rapidly & accurately by displaYing
HR RESP and oxygen parameter on same screen
Simultaneous Seven leads ECG display
| |
Temperature
(Skin or Intra-eavity)
| Units
| :C
or "F Measurement Range 0-50"C | | Resolution
| :0.1
·C | | Accuracy
| :
± 0.1 "C (excludes the sensor) | | Channel
| :
2-Channels | | Alarm
Range | 0-50
·C | | Respiration
Rate (RA·LL Impedance) | |
Method
| :
Impedance between RA-Ll | | Measurement
Range | :
Adult 0-120rpm Paed/Neonate: 0-l50, rpm | |
Accuracy
| +
2 rpm | | Alarm
Range | :
Adult : 0-120 rpm | | Apnea
Alarm | :
YES | | Apnea
Delay | :10-40
sec | |
| 2-Invasive
Blood Pressure (Optional) | |
Measurement Range: |
:ISO-300 mmHg | |
Channel |
: 2-Channels | |
Pressure Sensor |
: Sensitivity: 5uVNlrnmHg Impedance: +300-+10000 Ohms |
|
Pressure Name |
: ART. PA. CVP. RAP LAP, ICP | |
Resolution |
: 1 mmHg | |
Accuracy |
: ± 2 or ± 1 mmHg, whichever IS greater excludes sensor |
|
Alarm Range |
: -50-300 mmHg | | |
| Pulse
Rate | |
Measurement Range: |
: 18-254 BPM | |
Alarm Range : |
: 18-254 BPM | |
Accuracy |
: ± 3 BPM | |
Resolution : |
1 BPM | | |
| Electrical
| |
|
Power |
: 100-240 VAC. 50/60 HL | |
Battery |
. Rechargeable Li-ion, 2A!hr provides backup of lIpto 3 hrs |
| |
| Physical |
|
Dimension |
: 318 (W) x 152 (D) x 264 (1-1) mm : | |
Weight |
: -5 Kg | 
Big Digit Display
Big
digit display format makes reading parameters from a distance easy

Trend Display View
|

In-hospital applications
include emergency, pre-and post-operative care, ambulatory surgery, intermediate
care/step down units, labor and hospital-based special procedure areas.
- Advanced streamline outline design, compact,
lightweight, in-built handle makes it truly portable
- Can
be used to monitor adult, pediatric & neonate patient
- Upto
eight wave-channel display
- In-built
Rechargeable battery which provides a good backup lime
- 96
Hours graphical and tabular trends memory
- With
function of arrhythmia analysis, drug dose calculation, pacemaker detection and
ST segment analysis.
- User friendly
menu structure with easy to use rotary dial
- Protection
Against electro surgical interference and defibrillation
- Intelligent
audio and visual comprehensive alarm. Volume adjustment possible
- Central
monitoring system for long-distance monitoring, diagnosis, maintenance and software
upgrade possible
Standard
Configuration: ECG, HR, Sp02' ST Analysis, Resp, NIBP, Temp, Drug
Dose Calculation, built in battery Optional Configuration: Recorder,
2*IBP, CO2
Technical
Specifications:
| ECG
Input:
5 wires/3 wires ECG cable Lead
selection: I,II,III, aVR, aVL, aVF, V Gain
(mm/mV) : 1/4,1/2, 1, 2, 4 & auto Frequency response: Diagnosis 0.05 -
100Hz
Monitor 0.5
- 40 Hz
Surgery 1.0 - 25 Hz Sweep speed (mm/sec) : 12.5, 25, 50 Heart rate range:
10-300 BPM Heart
rate accuracy: ± 1 % or ± 1 BPM, which is greater ST segment deviation analysis:
-0.8 - +0.8mV ST
accuracy: ±0.02mV Heart
Rate Alarm: Yes Non-Invasive
Blood Pressure (Oscillographic) Measurement type: adult, pediatric, neonate
Measurement range: Systolic 30~270mmHg
Diastolic 10~220mmHg
Mean 20~235mmHg Accuracy:
± 2% or ± 3mmHg, which is greater Resolution:
1 mmHg Protection:
over pressure Pulse
Oximetry SpO2
measurement range: 0 -100 % Resolution:
1 % Pulse
measurement range: 25- 250BPM Accuracy:
± 2 %. 70~100% Resolution:
1 BPM Physical
Weight:
~ 4.5 Kg Size:
318mm x 264mm x 152mm

OxyCRG
Diagram Enables the doctor to relate information of patient rapidly &
accurately by displaying HR, RESP and oxygen parameter on same screen
 Simultaneous
Seven leads ECG display | Temperature
(Skin or Intra-Cavity) Measurement range: 0~500C Accuracy: ± 0.10C
Resolution:
0.10C Response
time: < 3 min. Channel:
1 Channel Respiration Rate (RA-LL Impedance) Measurement
range: 0 - 120 BPM Accuracy: ± 2 BPM Resolution:
1 BPM EtC02
(Optional Side Stream Type) Measurement range: 0%-13.1 % (0~99mmHg) Accuracy:
± 0.3%(< 5% Measurement value) <10%
(> 5% Measurement value) Response Time: 180 ms 2·lnvasive
Blood Pressure (Optional) Measurement range: -45~+300mmHg Channel:
2 channel Transducer
sensitivity: 5uV/V/mmHg, 2% Unit
display: kPa or mmHg selectable Environment
Operation
Storage
Temperature 0- 40·C -20 - 500C Relative humidity < 85%
< 93% Electrical
Input:
100~240 V AC, 50/60Hz Consumption:
< 70VA Battery
type: Lead-Acid Battery
Operation: ~ 2 hrs 
Big
Digit Display Big digit display format makes reading parameters from
a distance easy |
Next