Cardiac Care & ICU Equipments |
|

In-hospital applications
include emergency, pre-and post-operative care,
ambulatory surgery, intermediate care/step down
units, labor delivery and hospital-based special
procedure areas.
- Advanced streamline
outline design, compact, lightweight, in-built
handle makes it truly portable
- Can be used to
monitor adult, pediatric & neonate patient
- Upto seven wave-channel
display
- In-built Rechargeable
Li-ion battery which provides a good backup
time
- 72 Hours graphical
and tabular trends memory
- With function
of arrhythmia analysis, drug dose calculation,
pacemaker detection and ST segment analysis.
- User friendly
menu structure with easy to use rotary dial
- Protection Against
electro surgical interference and defibrillation
- Intelligent audio
and visual comprehensive alarm. Volume adjustment
possible
- Central monitoring
system to make long-distance monitoring, diagnosis,
maintenance and software upgrade possible
Technical
Specifications
|
ECG
Input: 5 wires/3 wires ECG cable
lead selection: I,II,III, aVR, aVl, aVF,
V
Gain (mm/mV): 1/4,1/2,1,2,4 &
auto
Frequency response: Diagnosis 0.05 - 100Hz
Monitor
0.5 - 35 Hz
Surgery
1.0 - 15 Hz
Sweep speed (mm/sec) : 12.5,25,50
Heart rate range: 15-300 BPM
Heart rate accuracy: ± 1 % or ± 1 BPM, which
is greater
ST segment deviation analysis: -2.0 ~ +2.0mV
ST accuracy: ±0.02mV
Heart Rate Alarm: Yes
Non-Invasive
Blood Pressure (Oscillographic)
Measurement type: adult, pediatric, neonate
Measurement range: Systolic 40~270mmHg
Diastolic
10~215mmHg
Mean 20~235mmHg
Accuracy: ± 2% or ± 5mmHg, which is greater
Resolution: 1 mmHg
Protection: over pressure
Pulse
Oximetry
SpO2
measurement range: 0 - 100 %
Resolution: 1%
Pulse
measurement range: 0- 254BPM
Accuracy: ± 2 %. 70~100%
Resolution:
1 BPM
|
2-Temperature·
Skin & Intra-Cavity (Optional)
Measurement range: 0~500C
Accuracy: ± 0.10C
Resolution: 0.10C
Response time: < 3 min.
Channel: 2 Channel
Respiration
Rate (Optional)
Method: RA-LL Impedance
Measurement range: 0 - 120 RPM
Accuracy: ± 2 RPM
Resolution: 1 RPM
Apnea Alarm : 10 ~ 40 Sec.
Electrical
Input: 100~250 V AC, 50/60Hz
Consumption: < 110VA
Battery type: Li-ion
Battery Operation: ~ 3 hrs
Physical
Weight: ~ 3.5 Kg
Size: 210mm x 235mm x 122mm
|
Standard
Configuration: ECG,
HR, Sp02, ST Analysis, NIBP, Drug Dose Calculation,
built in battery
Optional
Configuration: Temp *2, Resp, Recorder, 2*IBP,
CO2
|

In-hospital
applications include emergency, pre-and post-operative
care, ambulatory surgery, intermediate care/step
down units, labor and hospital-based special procedure
areas.
- Advanced streamline
outline design, compact, lightweight, in-built
handle makes it truly portable
- Can be used to
monitor adult, pediatric & neonate patient
- Upto eight wave-channel
display
- In-built Rechargeable
battery which provides a good backup lime
- 96 Hours graphical
and tabular trends memory
- With function
of arrhythmia analysis, drug dose calculation,
pacemaker detection and ST segment analysis.
- User friendly
menu structure with easy to use rotary dial
- Protection Against
electro surgical interference and defibrillation
- Intelligent audio
and visual comprehensive alarm. Volume adjustment
possible
- Central monitoring
system for long-distance monitoring, diagnosis,
maintenance and software upgrade possible
Standard
Configuration: ECG, HR, Sp02' ST
Analysis, Resp, NIBP, Temp, Drug Dose Calculation,
built in battery
Optional Configuration: Recorder, 2*IBP,
CO2
Technical
Specifications:
|
ECG
Input:
5 wires/3 wires ECG cable
Lead
selection: I,II,III, aVR, aVL, aVF, V
Gain
(mm/mV) : 1/4,1/2, 1, 2, 4 & auto
Frequency response: Diagnosis 0.05 - 100Hz
Monitor
0.5 - 40 Hz
Surgery 1.0 - 25 Hz
Sweep speed (mm/sec) : 12.5, 25, 50
Heart rate range: 10-300 BPM
Heart
rate accuracy: ± 1 % or ± 1 BPM, which is
greater
ST segment deviation analysis: -0.8 - +0.8mV
ST
accuracy: ±0.02mV
Heart
Rate Alarm: Yes
Non-Invasive
Blood Pressure (Oscillographic)
Measurement type: adult, pediatric, neonate
Measurement range: Systolic 30~270mmHg
Diastolic
10~220mmHg
Mean 20~235mmHg
Accuracy:
± 2% or ± 3mmHg, which is greater
Resolution:
1 mmHg
Protection:
over pressure
Pulse
Oximetry
SpO2
measurement range: 0 -100 %
Resolution:
1 %
Pulse
measurement range: 25- 250BPM
Accuracy:
± 2 %. 70~100%
Resolution:
1 BPM
Physical
Weight:
~ 4.5 Kg
Size:
318mm x 264mm x 152mm

OxyCRG
Diagram
Enables the doctor to relate information
of patient rapidly & accurately by displaying
HR, RESP and oxygen parameter on same screen

Simultaneous
Seven leads ECG display
|
Temperature
(Skin or Intra-Cavity) Measurement range:
0~500C Accuracy: ± 0.10C
Resolution:
0.10C
Response
time: < 3 min.
Channel:
1 Channel
Respiration Rate (RA-LL Impedance)
Measurement range: 0 - 120 BPM Accuracy: ±
2 BPM
Resolution:
1 BPM
EtC02
(Optional Side Stream Type) Measurement
range: 0%-13.1 % (0~99mmHg)
Accuracy: ± 0.3%(< 5% Measurement value)
<10%
(> 5% Measurement value) Response Time:
180 ms
2·lnvasive
Blood Pressure (Optional) Measurement
range: -45~+300mmHg
Channel:
2 channel
Transducer
sensitivity: 5uV/V/mmHg, 2%
Unit
display: kPa or mmHg selectable
Environment
Operation
Storage
Temperature 0- 40·C -20 - 500C
Relative humidity < 85% <
93%
Electrical
Input:
100~240 V AC, 50/60Hz
Consumption:
< 70VA
Battery
type: Lead-Acid
Battery
Operation: ~ 2 hrs

Big
Digit Display
Big digit display format makes reading
parameters from a distance easy
|
Multipara
Monitor Model MPM121
Also available in 12.1" Screen Size

In-hospital
applications include emergency pre-and post-operative
care, ambulatory surgery, intermediate care/step
down units, labor and delivery, and hospital-based
special procedure areas.
Suitable
for use in physicians’ offices, clinics, outpatient
surgical centers, extended care facilities and other
patient care areas, which require affordable monitoring
- Advanced streamline
outline design, portable, compact, lightweight
- 12.1" color TFT
LCD screen with upto eight wave-channel display
- Brightness Control
to make patient comfortable at night
- In-built Rechargeable
battery which provides a good backup time
- 360 Hours graphical
and tabular trends memory
- With function
of venous punch, drug dose calculation, pacemaker
detection and ST segment analysis.
- Operating menu
with multi-language interface selection: English,
Spanish, Portuguese, Chinese, French, etc.
- User friendly
menu structure with easy to use rotary dial
- Protection Against
electrosurgical interference and defibrillation
- Intelligent audio
and visual comprehensive alarm
- WAN communication
function to network with central monitoring
system and make long-distance monitoring, diagnosis,
maintenance and software upgrade possible
- Optional built-in
wireless networking function
- Supports external
printer to record graph and trend
Standard
Configuration: ECG, HR, PR, SpO2, ST Analysis,
Resp * 2, NIBP (with venous punch), Temp *2, Drug
Dose Calculation, built in Li-ion Battery.
Optional
Configuration: Recorder, CO2.
Technical Specifications:
|
Non-invasive
blood pressure
|
Temperature
|
|
Measurement
type: adult, pediatric, neonate
|
Measurement
range: 25 – 45 ° C
|
|
Measurement
range:
|
Systolic
Diastolic
Mean
|
4.0
- 37.0 kPa
1.3 - 33.0 kPa
2.6 - 35.0 kPa
|
Accuracy:
± 0.1 ° C
|
|
Accuracy
: ± 0.4 kPa or 5 %
|
Resolution:
0.1 ° C
|
|
Resolution:
0.1 kPa
|
response
time: ≤ 3 min.
|
|
Protection:
over pressure
|
|
|
|
|
ECG
|
Respiration
Rate
|
|
Input:
5 wires ECG cable
|
Measurement
range: 0 - 120 BPM
|
|
Lead
section: I,II,III/ aVR, aVL, aVF/ V
|
Accuracy:
± 1 BPM or 5 %
|
|
Gain
(mm/mV): ¼,1/2, 1, 2, 4
|
Resolution:
1 BPM
|
|
Frequency
response:
|
Diagnosis
Monitor
Surgery
|
0.05
- 100Hz
0.5 - 40 Hz
1.0 - 25 Hz
|
CO2
|
|
Sweep
speed (mm/sec): 12.5, 25, 50
|
Measurement
range: 0 %-10.0 % (ins CO2 and et CO2)
|
|
Heart
rate range: 30-254 BPM
|
Accuracy:
± 2 %(< 5 % Measurement
value)
|
|
Heart
rate accuracy: ± 1 %
|
10
% (> 5 % Measurement
value)
|
|
ST
segment deviation analysis: -0.8-+0.8mV
|
Pulse
oximetry
|
|
ST
accuracy: ±0.02mv
|
SPO2
measurement range: 0 - 100 %
|
|
Invasive
blood pressure
|
Resolution:
1 %
|
|
Measurement
range: 0~40kPa (0~300mmHg)
|
pulse
measurement range: 30- 254BPM
|
|
Channel:
2 channel
|
Accuracy:
± 2 %
|
|
Transducer
sensitivity: 5MV/V/mmHg
|
Resolution:
1 BPM
|
|
Unit
display: kPa or mmHg selectable
|
|
|
|
Environment
|
|
Power
requirements
|
|
Operation
|
Storage
|
|
Input:
98~260 V AC, 50/60Hz
|
Temperature
|
0.5
- 40 ° C
|
-20
- 50 ° C
|
|
Consumption:
≤ 80VA
|
Relative
humidity
|
≤
80%
|
≤
90%
|
<<BACK
|
| | 
|