In-hospital applications include emergency, pre-and post-operative care, ambulatory surgery, intermediate care/step down units, labor delivery and hospital-based special procedure areas.

  • Advanced streamline outline design, compact, lightweight, in-built handle makes it truly portable
  • Can be used to monitor adult, pediatric & neonate patient
  • Upto seven wave-channel display
  • In-built Rechargeable Li-ion battery which provides a good backup time
  • 72 Hours graphical and tabular trends memory
  • With function of arrhythmia analysis, drug dose calculation, pacemaker detection and ST segment analysis.
  • User friendly menu structure with easy to use rotary dial
  • Protection Against electro surgical interference and defibrillation
  • Intelligent audio and visual comprehensive alarm. Volume adjustment possible
  • Central monitoring system to make long-distance monitoring, diagnosis, maintenance and software upgrade possible
Technical Specifications

ECG
Input: 5 wires/3 wires ECG cable
lead selection: I,II,III, aVR, aVl, aVF, V
Gain (mm/mV): 1/4,1/2,1,2,4 & auto
Frequency response: Diagnosis 0.05 - 100Hz
                                    Monitor 0.5 - 35 Hz
                                    Surgery 1.0 - 15 Hz
Sweep speed (mm/sec) : 12.5,25,50
Heart rate range: 15-300 BPM
Heart rate accuracy: ± 1 % or ± 1 BPM, which is greater
ST segment deviation analysis: -2.0 ~ +2.0mV
ST accuracy: ±0.02mV
Heart Rate Alarm: Yes

Non-Invasive Blood Pressure (Oscillographic)
Measurement type: adult, pediatric, neonate Measurement range: Systolic 40~270mmHg
                                    Diastolic 10~215mmHg
                                    Mean 20~235mmHg
Accuracy: ± 2% or ± 5mmHg, which is greater Resolution: 1 mmHg
Protection: over pressure

Pulse Oximetry

SpO2 measurement range: 0 - 100 %
Resolution: 1%
Pulse measurement range: 0- 254BPM
Accuracy: ± 2 %. 70~100%
Resolution: 1 BPM

2-Temperature· Skin & Intra-Cavity (Optional)
Measurement range: 0~500C
Accuracy: ± 0.10C
Resolution: 0.10C
Response time: < 3 min.
Channel: 2 Channel

Respiration Rate (Optional)
Method: RA-LL Impedance
Measurement range: 0 - 120 RPM
Accuracy: ± 2 RPM
Resolution: 1 RPM
Apnea Alarm : 10 ~ 40 Sec.

Electrical
Input: 100~250 V AC, 50/60Hz
Consumption: < 110VA
Battery type: Li-ion
Battery Operation: ~ 3 hrs

Physical
Weight: ~ 3.5 Kg
Size: 210mm x 235mm x 122mm

Standard Configuration: ECG, HR, Sp02, ST Analysis, NIBP, Drug Dose Calculation, built in battery

Optional Configuration: Temp *2, Resp, Recorder, 2*IBP, CO2



In-hospital applications include emergency, pre-and post-operative care, ambulatory surgery, intermediate care/step down units, labor and hospital-based special procedure areas.

  • Advanced streamline outline design, compact, lightweight, in-built handle makes it truly portable
  • Can be used to monitor adult, pediatric & neonate patient
  • Upto eight wave-channel display
  • In-built Rechargeable battery which provides a good backup lime
  • 96 Hours graphical and tabular trends memory
  • With function of arrhythmia analysis, drug dose calculation, pacemaker detection and ST segment analysis.
  • User friendly menu structure with easy to use rotary dial
  • Protection Against electro surgical interference and defibrillation
  • Intelligent audio and visual comprehensive alarm. Volume adjustment possible
  • Central monitoring system for long-distance monitoring, diagnosis, maintenance and software upgrade possible

Standard Configuration: ECG, HR, Sp02' ST Analysis, Resp, NIBP, Temp, Drug Dose Calculation, built in battery
Optional Configuration: Recorder, 2*IBP, CO2


Technical Specifications:

ECG
Input: 5 wires/3 wires ECG cable
Lead selection: I,II,III, aVR, aVL, aVF, V
Gain (mm/mV) : 1/4,1/2, 1, 2, 4 & auto
Frequency response: Diagnosis 0.05 - 100Hz
                                   
Monitor 0.5 - 40 Hz
                                    Surgery 1.0 - 25 Hz
Sweep speed (mm/sec) : 12.5, 25, 50
Heart rate range: 10-300 BPM
Heart rate accuracy: ± 1 % or ± 1 BPM, which is greater
ST segment deviation analysis: -0.8 - +0.8mV
ST accuracy: ±0.02mV
Heart Rate Alarm: Yes

Non-Invasive Blood Pressure (Oscillographic)
Measurement type: adult, pediatric, neonate
Measurement range: Systolic 30~270mmHg

                           Diastolic 10~220mmHg
                                    Mean 20~235mmHg

Accuracy: ± 2% or ± 3mmHg, which is greater
Resolution: 1 mmHg
Protection: over pressure

Pulse Oximetry
SpO2 measurement range: 0 -100 %
Resolution: 1 %
Pulse measurement range: 25- 250BPM
Accuracy: ± 2 %. 70~100%
Resolution: 1 BPM

Physical
Weight: ~ 4.5 Kg
Size: 318mm x 264mm x 152mm



OxyCRG Diagram
Enables the doctor to relate information of patient rapidly & accurately by displaying HR, RESP and oxygen parameter on same screen


Simultaneous Seven leads ECG display

Temperature (Skin or Intra-Cavity) Measurement range: 0~500C Accuracy: ± 0.10C
Resolution: 0.10C
Response time: < 3 min.
Channel: 1 Channel


Respiration Rate (RA-LL Impedance)
Measurement range: 0 - 120 BPM Accuracy: ± 2 BPM
Resolution: 1 BPM

EtC02 (Optional Side Stream Type) Measurement range: 0%-13.1 % (0~99mmHg)
Accuracy: ± 0.3%(< 5% Measurement value)

<10% (> 5% Measurement value) Response Time: 180 ms

 

2·lnvasive Blood Pressure (Optional) Measurement range: -45~+300mmHg
Channel: 2 channel
Transducer sensitivity: 5uV/V/mmHg, 2%
Unit display: kPa or mmHg selectable

Environment

Operation Storage
Temperature 0- 40·C -20 - 500C
Relative humidity < 85% < 93%

Electrical
Input: 100~240 V AC, 50/60Hz
Consumption: < 70VA
Battery type: Lead-Acid
Battery Operation: ~ 2 hrs





Big Digit Display
Big digit display format makes reading parameters from a distance eas
y



Multipara Monitor Model MPM121
Also available in 12.1" Screen Size

In-hospital applications include emergency pre-and post-operative care, ambulatory surgery, intermediate care/step down units, labor and delivery, and hospital-based special procedure areas.

Suitable for use in physicians’ offices, clinics, outpatient surgical centers, extended care facilities and other patient care areas, which require affordable monitoring

  • Advanced streamline outline design, portable, compact, lightweight
  • 12.1" color TFT LCD screen with upto eight wave-channel display
  • Brightness Control to make patient comfortable at night
  • In-built Rechargeable battery which provides a good backup time
  • 360 Hours graphical and tabular trends memory
  • With function of venous punch, drug dose calculation, pacemaker detection and ST segment analysis.
  • Operating menu with multi-language interface selection: English, Spanish, Portuguese, Chinese, French, etc.
  • User friendly menu structure with easy to use rotary dial
  • Protection Against electrosurgical interference and defibrillation
  • Intelligent audio and visual comprehensive alarm
  • WAN communication function to network with central monitoring system and make long-distance monitoring, diagnosis, maintenance and software upgrade possible
  • Optional built-in wireless networking function
  • Supports external printer to record graph and trend

Standard Configuration: ECG, HR, PR, SpO2, ST Analysis, Resp * 2, NIBP (with venous punch), Temp *2, Drug Dose Calculation, built in Li-ion Battery.

Optional Configuration: Recorder, CO2.

 

Technical Specifications:

Non-invasive blood pressure

Temperature

Measurement type: adult, pediatric, neonate

 Measurement range: 25 – 45 ° C

Measurement range:

Systolic
Diastolic
Mean

4.0 - 37.0 kPa
1.3 - 33.0 kPa
2.6 - 35.0 kPa

 Accuracy: ± 0.1 ° C

Accuracy : ± 0.4 kPa or 5 % 

Resolution: 0.1 ° C

Resolution: 0.1 kPa

response time: ≤ 3 min.

Protection: over pressure

ECG

Respiration Rate

Input: 5 wires ECG cable

Measurement range: 0 - 120 BPM

Lead section: I,II,III/ aVR, aVL, aVF/ V

Accuracy: ± 1 BPM or 5 %

Gain (mm/mV): ¼,1/2, 1, 2, 4

Resolution: 1 BPM

Frequency response:

Diagnosis
Monitor
Surgery

0.05 - 100Hz
0.5 - 40 Hz
1.0 - 25 Hz

CO2

Sweep speed (mm/sec): 12.5, 25, 50

Measurement range: 0 %-10.0 % (ins CO2 and et CO2)

Heart rate range: 30-254 BPM

Accuracy: ± 2 %(< 5 % Measurement value)

Heart rate accuracy: ± 1 %

10 % (> 5 % Measurement value)

ST segment deviation analysis: -0.8-+0.8mV

Pulse oximetry

ST accuracy: ±0.02mv

SPO2 measurement range: 0 - 100 %

Invasive blood pressure

Resolution: 1 %

Measurement range: 0~40kPa (0~300mmHg)

pulse measurement range: 30- 254BPM

Channel: 2 channel

Accuracy: ± 2 %

Transducer sensitivity: 5MV/V/mmHg

Resolution: 1 BPM

Unit display: kPa or mmHg selectable

Environment

Power requirements

Operation

Storage

Input: 98~260 V AC, 50/60Hz

Temperature

0.5 - 40 ° C

-20 - 50 ° C

Consumption: ≤ 80VA

Relative humidity

≤ 80%

≤ 90%

 

 

 

<<BACK

 



About us                 |             Product Range                  |            Enquiry Form                    |             Contact            |             Home