    FEATURES


    14.4cm (5.7") STN Blue LCD with backlit display

    Non fade soild traces of graphical ECG waveform on the LCD Screen

    Real time ECG on the fist channel and frozen / delayed ECG on the Second channel

    Auto ECG freeze on alarm condition

    1mV calibration pulse indicator

    LED blinking for every delected QRS complex

    Standared frequency responese and high common mode rejection ratio

    Built-in rechargeable battery

     

    TECHNICAL SPECIFICATIONS 

    GENERAL

    Display

    STN Blue LCD with CCFL back light 

     

    Display area

    120mm x 90mm

     

    Resolution

    320 x 240 pixels

     

    Writing speed

    25mm/sec

     

    Alarms

    Visual (Bell) and Audio (Buzzer beep) 

     

    Freeze

    On demand or automatic on alarm Frozen display on channel 2 

     

    Key board

    Flat touch membrane operation keys

     

    CNS back panel connector

    9 Pin D-sub connector 0.5VjmV at 10mmjmV gain 

     

    Dimensions

    235 x 166 x 140mm 

     

    Weight

    1.5 kgs 

    POWER SUPPLY

    AC - DC Adapter 

    15V DC/1.5A 

     

    AC mains

    100 - 264 V AC

     

    Line indicator

    Green LED 

     

    Power consumption 

    < 18VA

     

    Battery 

    Rechargeable NiMH 9.6V 1400mAh 

     

    Battery backup duration

    180 minutes (min) 

     

    Battery low indicator 

    Red LED 

     

    ECG display

    4 seconds real time ECG on channel 2

    ECG 

    Leads

    User selectable I, II, III 

     

    Lead indicator 

    Yellow LEDs 

     

    Heart rate range 

    30 - 250 BPM 

     

    Heat rate accuracy

    _+2BPM or 2% whichever is higher

     

    Low alarm limit range

    30 to 240 BPM

     

    High alarm limit range

    40 to 250 BPM

     

    Bandwidth

    0.5 Hz to 35 Hz 

     

    User gain settings for 10mm display


    lmV, 0.66mV, 0.5mV & 2mV display
     

     

    CMRR

    > 90dB @ 50Hz 

     

    Input impedance 

    > 205M ohms 

     

    Leakage current 

    < 20uA 

     

    Patient isolation

    > 2000 V AC @ 50Hz for 1 minute 

     

    Reset recovery 

    Automatic return of waveform within 0.5 seconds after defibrillation or electrical overload 

     

    Defibrillator protected

    Yes 

     

    Patient safety 

    Class 1 CF 

    ENVIRONMENTAL 

    Operating temperature 

    10° to 40° C 

     

    Relative humidity

    10 to 90% non condensing

    STANDARD ACCESSORIES 

    ECG Patient cable

    1 no. 

     

    Disposable electrodes

    30 nos. 

     

    AC - DC adapter (AD-004)

    1 no. 

     

    Accessory bag 

    1 no. 

     

    User manual 

    1 no. 

    OPTIONAL ACCESSORY 

    Wall mountable stand

    1 no

    * Due to constant upgradation, specifications and prices are subject to change without prior notice.