Power
supply: 220V±10%. 5OHz±1Hz
Internal battery: DC12V (over for four
hours)
Power consumption: <50W
|
Patient
type: Adult, Pediatric
Driven
mode: Pneumatically driven electronically
controlled
Mode: Semi-open. Semi-closed or closed
system
|
|
Respiration
Pattern: IPPV, SIMV. SIGH. MANUAL. Display:
5.5" SNT screen
Waveform:
P-t, V-t
Tidal
volume: 0. 50ml-1600ml
Respiratory frequency: 1 to 65 BPM
I: E ratio: 8:1-1:8
Pressure
high alarm limit: 20-60cmH20
Pressure low alarm Limit: 0-20cmH20
Monitor
Parameters: tidal volume. MV, Ppeak. PEEP,
respiratory frequency, I:E ratio.
|
Alarm
parameters:
pressure high alarm
pressure low alarm
O2 lost alarm
AC
power lost alarm
Alarm silence function
Safety
valve pressure limit: 6Kpa
Good
air tightness, high reliability, and minor
respiration impede ance
|
|
Circle
absorber: compact circle absorber
Bellow: 0-1600ml
Flow
meter: O2,N20 four-tubes
flow meter
O2 tube: 0L/min-1L/min··10L/min
N2Otube: 0L/min-1L/min-10L/min
O2
and N20 Linkage Device
Adjust
O2 and N2O in proportion(O2
concentration is not lower than25%)
|
Oxygen
FLUSH: 35L/min-7SUmin Gas source pressure:
O2:
0.27MPa-O.55MPa
N20:
0.27MPa-O.55MPa
Gauge:
two gauges
Anesthetic
vaporizer: one vaporizer, two vaporizer
is option
|