Ventilators

  • Lightweight, durable and simple to use
  • Ventilation volume and ventilation frequency can be adjusted in succession and separately.
  • Manometer to indicate airway pressure
  • Automatic pressure relief valve for airway pressure more than 6KPa
  • Audible low input pressure alarm and over input pressure protection
  • Indicator for inspiratory & expiratory phase differentiation

Specifications:

VENTILATOR PARAMETERS

RANGE

Mode

Control (IPPV)

Tidal Volume

100~1000 ml

Breath Rate

10, 15, 20, 25, 30 bpm

%02

45%

I: E

2:1, 1:1, 1:1.5, 1:2, 1:2.5, 1:3

   

ALARMS AND INDICATORS

 

Over Pressure Relief

< 60 hpa

Low Battery Alarm, Power & Charging Indicator

 

ELECTRICAL SPECIFICATIONS INPUT POWER SUPPLY
DC Voltage: 13 V
Power: 25 VA
 
CHARGER
AC Voltage: 100 ~ 240 V, 50/60 Hz
Power: 25 VA
 
 
PNEUMATIC SPECIFICATIONS
Oxygen Inlet
Pressurized Oxygen: 0.40 Mpa, 60 L/min.
 
PHYSICAL
355mm (W) x 250mm (D) x 140mm (H)
Weight: 3 Kg


 

EMC Portable Emergency Ventilator

 

  • Respiratory Rate and Minute Ventilation Volume can be independently regulated at random to meet various patients' requirements.
  • Manual Ventilation Mode, PEEP
  • The airway pressure can be shown directly and reduced automatically.
  • Air-Oxygen Mixer
  • Low-pressure Alarm
  • Assistant Ventilation Mode can support oxygen for patients’ inspiratory
    effort by automatic flow trigger

Standard configuration
Ventilator ( including air-oxygen mixer, automatic airway pressure relief valve ), small oxygen cylinder, portable bag, laryngoscope, phlegm suction pot, humidifier, airway opener, extension line, oxygen refill fitting, wrench, face mask, nasal oxygen canula, endotracheal tube.

 Physical parameters

 

Weight of Main Device

0.70kg 

Volume of Main Device

178×46×68mm

Weight of Gas Cylinder

2.5kg

Volume of Gas Cylinder

2L

Main technical indicator

Breathing Mode

IPPV Mode; A/C Ventilation; Manual Ventilation; PEEP

Operating Pressure

 0.35--0.40 MPa

Minute Ventilation Volume

4--25L/min(Adjustable in succession)

Respiratory Rate

10-60 times/min  (Adjustable in succession)
Seven Indication: 10,15,20,30,40,50,60 times/min

I /E Ratio

1:2

Oxygen Concentration

≤60%

Inspiratory Trigger

≤-1Kpa  (±50Pa)

Low Pressure Alarm

≤0.25 MPa

High Pressure Pause

0.9-1.1 MPa

Reverse Flow Pressure of Airway

≤6 KPa

Noise of Main Device

65dB

Supply Flow of Gas Cylinder

60 L/min

Continuous Working Time of Cylinder

Around 90min

Oxygen Storage Pressure for Cylinder

14.7 MPa

Oxygen Consumption

Adult ≤ 2 Cylinders/24h
Cylinder Specification: 12250 Kpa/40L

Protective Function

Airway Pressure Display

Visual

Low Pressure Alarm

Warning when supplied air pressure drops below 2.6bar

Airway Pressure Release Valve

Reducing pressure automatically when airway pressure exceeds 50cmH2O

High Pressure Protection

When supplied air pressure is over the working pressure, the ventilator will stop working following audible alarm voice.

 (Respiratory Rate, Minute Ventilation Volume, I:E Ratio, Oxygen Concentration, Inspiratory Trigger are subject to a tolerance of ±15%)


 

EMD Portable Emergency Ventilator

  • Lightweight, Durable &Simple to use.
  • Ventilation frequency & tidal volume can be adjusted in succession & separately
  • Airway Pressure is shown directly on the airway manometer.
  • Automatic Airway Pressure release on reaching 6KPa
  • Unique Air Oxygen Blender prolongs the oxygen cylinder to help during transportation.
  • Audible Alarm for Low Input Pressure
  • Protection for Over Input Pressure

Standard configuration
Ventilator ( including air-oxygen mixer, automatic airway pressure relief valve ), small oxygen cylinder, portable bag, laryngoscope, phlegm suction pot, humidifier, airway opener, extension line, oxygen refill fitting, wrench, face mask, nasal oxygen canula, endotracheal tube.

Physical parameters

Weight of main device

 1.03kg 

Volume of main device

190×82×65mm

Weight of gas cylinder

2.5kg


Technical Specifications

Breathing Mode

IPPV  Mode

Ventilation Volume

2-24 L/min 

Respiratory Rate/Ventilation Frequency

10-30 bpm (Adjustable in succession) 

I /E Ratio

1:2

Reverse Flow Pressure of Airway

≤ 6kPa  (regular)

Demand Trigger Breathing Sensitivity

≤ -1kPa  

Low Pressure Alarm

≤ 0.25MPa  

Volume of the Gas Cylinder

2.8L (Compressed gas)

Oxygen Storage Pressure for Cylinder

20MPa

Continuous Working Time of Cylinder

Around 90min

Oxygen Concentration

≤60%

 


About us                 |             Product Range                  |            Enquiry Form                    |             Contact            |             Home